4.4 Article

Fundamental uncertainty analysis of flowrate measurement using the ultrasonic Doppler velocity profile method

Journal

FLOW MEASUREMENT AND INSTRUMENTATION
Volume 33, Issue -, Pages 202-211

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.flowmeasinst.2013.07.004

Keywords

Flowrate; Velocity profile; Calibration; Uncertainty; Doppler; Ultrasonic

Ask authors/readers for more resources

The present paper describes a fundamental uncertainty analysis for a flowrate measurement in a pipe using an ultrasonic Doppler velocity profile method and an evaluation of the estimated uncertainty by an actual flow calibration. The uncertainties are estimated for internal factors originating from the measurement equipment; UVP provided by Met-Flow sa. and external factors depending on on-site measurements, such as the inclination angle of the ultrasonic transducer. The relative expanded uncertainty due to internal factors is estimated to be 0.34% with a coverage factor of 2. The relative external uncertainty including external factors is estimated to be from 0.42% to 2.13% depends on the inclination angle of the transducer. The results of the actual flow calibration under the same condition as the uncertainty analysis are within the range of uncertainty considering the internal factors. (C) 2013 Published by Elsevier Ltd.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available