4.7 Article

Fully Automated Versus Standard Tracking of Left Ventricular Ejection Fraction and Longitudinal Strain The FAST-EFs Multicenter Study

Journal

JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY
Volume 66, Issue 13, Pages 1456-1466

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jacc.2015.07.052

Keywords

agreement; automated function; echocardiography; observer variation; software

Funding

  1. TomTec Imaging Systems

Ask authors/readers for more resources

BACKGROUND Echocardiographic determination of ejection fraction (EF) by manual tracing of endocardial borders is time consuming and operator dependent, whereas visual assessment is inherently subjective. OBJECTIVES This study tested the hypothesis that a novel, fully automated software using machine learning-enabled image analysis will provide rapid, reproducible measurements of left ventricular volumes and EF, as well as average biplane longitudinal strain (LS). METHODS For a total of 255 patients in sinus rhythm, apical 4- and 2-chamber views were collected from 4 centers that assessed EF using both visual estimation and manual tracing (biplane Simpson's method). In addition, datasets were saved in a centralized database, and machine learning-enabled software (AutoLV, TomTec-Arena 1.2, TomTec Imaging Systems, Unterschleissheim, Germany) was applied for fully automated EF and LS measurements. A reference center reanalyzed all datasets (by visual estimation and manual tracking), along with manual LS determinations. RESULTS AutoLV measurements were feasible in 98% of studies, and the average analysis time was 8 +/- 1 s/patient. Interclass correlation coefficients and Bland-Altman analysis revealed good agreements among automated EF, local center manual tracking, and reference center manual tracking, but not for visual EF assessments. Similarly, automated and manual LS measurements obtained at the reference center showed good agreement. Intraobserver variability was higher for visual EF than for manual EF or manual LS, whereas interobserver variability was higher for both visual and manual EF, but not different for LS. Automated EF and LS had no variability. CONCLUSIONS Fully automated analysis of echocardiography images provides rapid and reproducible assessment of left ventricular EF and LS. (C) 2015 by the American College of Cardiology Foundation.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available