The Best Conditions for Parallel Stenting During EVAR: An In Vitro Study

Journal

Publisher

W B SAUNDERS CO LTD
DOI: 10.1016/j.ejvs.2012.08.007

Keywords

Aortic aneurysm, Abdominal; Renal artery; Endovascular procedures/methods; Vascular surgical procedures/instrumentation; Stents; Models, Cardiovascular

Funding

  1. Medtronic Inc.
  2. W.L. Gore Associates

Ask authors/readers for more resources

Objectives: The aim of this study is to identify which endograft, and to what degree of oversizing, in combination with what type of parallel stent, may result in the most adequate fit in a juxtarenal abdominal aneurysmal neck when using a parallel-stent technique. Materials/Methods: In-vitro silicon aneurysmal neck models of different diameters, with one side-branch, were constructed. Two different endografts (Medtronic-Endurant Abdominal Stent Graft and Gore-Excluder abdominal aortic aneurysm Endoprosthesis; three diameters each), and two stents (self-expanding Gore Viabahn Endoprosthesis and balloon-expandable Atrium Advanta V12; 6-mm diameter) were tested, applying three endograft-oversizing degrees (15%, 30% and 40%). After remodelling using the kissing-balloon technique at 37 degrees C, the 36 endograft-stent-oversizing combinations were scanned by computed tomography (CT). The size of the results in gutters, parallel-stent compression and main stent-graft infolding were recorded. Results: Increasing oversizing (15%, 30% and 40%) significantly decreased gutter areas (11.5, 6.2, 4.3 mm(2), P < 0.001); nevertheless, main endograft infoIding of most 40%-oversized stent grafts was detected, particularly with Excluder devices. Lower stent compression, but wider gutters, were observed with the Excluder when compared to Endurant stent grafts, and with V12 when compared to Viabahn parallel stents. The Endurant-Viabahn combination resulted in maximum stent compression (35%). Conclusions: Better endograft-stent apposition was achieved when using 30% endograft oversizing. Lower stent compression, but wider gutters, were observed with the Excluder stent-graft and V12 parallel stent, achieving maximum stent compression with the Endurant-Viabahn combination. (c) 2012 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available