4.5 Article

Laryngeal pacing in minipigs: in vivo test of a new minimal invasive transcricoidal electrode insertion method for functional electrical stimulation of the PCA

Journal

EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY
Volume 270, Issue 1, Pages 225-231

Publisher

SPRINGER
DOI: 10.1007/s00405-012-2141-1

Keywords

Minipig; Larynx; Bilateral recurrent laryngeal nerve paralysis; Functional electric stimulation

Funding

  1. Med-El Medical Electronics Innsbruck

Ask authors/readers for more resources

Functional electrical stimulation (FES) of the posterior cricoarytenoid muscle (PCA) to restore respiratory function of the larynx may become an option for the treatment of bilateral recurrent laryngeal nerve paralysis (RLNP) in the near future. The feasibility of this has been shown in several animal trials and in a human pilot study. The common open surgical inferolateral approach for electrode insertion into the PCA for FES has a risk of damaging the recurrent laryngeal nerve (RLN) and may result in postoperative swelling and scaring of the larynx. Therefore, a minimal invasive electrode insertion technique is needed. A new miniaturized bipolar spiral tip electrode and a new electrical stimulatable insertion needle were tested in a short-term trial for an endoscopically guided and functionally controlled transcricoidal electrode insertion in eight Gottingen minipigs with bilateral normal RLN function. The feasibility of this technique was evaluated and the achieved positions of the electrodes in the PCA were analyzed using intraoperative stimulation threshold data and 3D-CT reconstructions. In seven cases it was possible to place two well-performing electrodes into the PCA. They were positioned one on either side. In one animal no functioning electrode position could be achieved because the PCA was missed. Thresholds of the electrode tips varied between 0.2 and 2.5 mA (mean 0.71 mA). In any case maximal glottal opening could be reached before adductors were co-activated. The majority of electrodes were placed into the central lower part of the PCA with no apparent correlation between threshold and electrode position. Surgical trauma might be further reduced by using endoscopy via a laryngeal mask avoiding the temporary tracheostomy used in this trial. If the implanted electrodes remain stable in long-term tests, we suggest that this method could soon be transferred into human application.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available