4.5 Article

An integrated on-chip sirtuin assay

Journal

ELECTROPHORESIS
Volume 31, Issue 19, Pages 3263-3267

Publisher

WILEY
DOI: 10.1002/elps.201000220

Keywords

High-throughput screening; Inhibitor screening; Lab-on-a-chip; Microchip; Sirtuin

Ask authors/readers for more resources

A microchip-based assay to monitor the conversion of peptide substrates by human recombinant sirtuin 1 (hSIRT1) is presented. For this purpose a fused silica microchip consisting of a microfluidic separation structure with an integrated serpentine micromixer has been used. As substrate for the assay, we used a 9-fluorenylmethoxycarbonyl (Fmoc)-labeled tetrapeptide derived from the amino acid sequence of p53, a known substrate of hSIRT1. The Fmoc group at the N-terminus resulting from solid-phase peptide synthesis enabled deep UV laser-induced fluorescence detection with excitation at 266 nm. The enzymatic reaction of 0.1 U/mu L hSIRT1 was carried out within the serpentine micromixer using a 400 mu M solution of the peptide in buffer. In order to reduce protein adsorption, the reaction channel was dynamically coated with hydroxypropylmethyl cellulose. The substrate and the deacetylated product were separated by microchip electrophoresis on the same chip. The approach was successfully utilized to screen various SIRT inhibitors.