4.3 Article

Tip-sensitive fibre-optic Bragg grating ultrasonic hydrophone for measuring high-intensity focused ultrasound fields

Journal

ELECTRONICS LETTERS
Volume 50, Issue 9, Pages 649-+

Publisher

INST ENGINEERING TECHNOLOGY-IET
DOI: 10.1049/el.2013.3961

Keywords

-

Funding

  1. National Natural Science Foundation of China [81127901]

Ask authors/readers for more resources

A tip-sensitive fibre-optic Bragg grating ultrasonic hydrophone (TSFBGUH) with high spatial resolution for measuring high-intensity focused ultrasound (HIFU) fields is reported. When measuring a HIFU field, the sensitive axis of the TSFBGUH with a flexible fibre-optic sensor holder should be in parallel with the acoustic axis of the HIFU transducer to ensure the measured ultrasound impinges on the tip of the TSFBGUH. The TSFBGUH system has been established and tested. The experimental results show that the acoustic pressure sensitivity of the TSFBGUH system is about 31.3 mV/MPa within the measurement range of 10 MHz and the noise equivalent pressure is about 10.4 kPa.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available