4.3 Article

Development of capacitance measurement system for human serum albumin detection

Journal

ELECTRONICS LETTERS
Volume 46, Issue 10, Pages 678-679

Publisher

INST ENGINEERING TECHNOLOGY-IET
DOI: 10.1049/el.2010.0234

Keywords

-

Ask authors/readers for more resources

The use of interdigitated electrodes and binding with Cibacron blue F3GA by self-assembly technology, which has high affinity for human serum albumin (HSA), and serves as a receptor for HSA detection, is reported. In preliminary tests, the interdigitated chip measured the impedance of HSA by electrochemical impedance spectroscopy and this result indicated the characteristic frequency of impedance change below 1.1 kHz. However, a portable capacitance measurement system made using a standard operational amplifier circuit and AC excitation was applied at 1 kHz. The detection result showed a significant linear correlation between HSA concentration (0.1 to 1 mg/ml) and output voltage. The regression equation was Y = 17.91 + 323.72X with R(2) = 0.999. This reported study has successfully demonstrated that the capacitive sensor has great potential for applications in HSA detection.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available