Journal
ELECTROCHIMICA ACTA
Volume 56, Issue 27, Pages 9883-9886Publisher
PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.electacta.2011.08.065
Keywords
Electrochemistry; Sensor; Glass pH electrode; Non-equilibrium; Sensitivity
Categories
Funding
- National Natural Science Foundation of China [20775050, 20927007]
Ask authors/readers for more resources
In this communication, a new four-electrode measurement system has been developed to improve the analytical performance of a glass pH electrode. The sensitivity of detection is greatly enhanced and the working range is widened. In this new configuration, the potentiometric response slope of the pH electrode, dE/dpH, could be as high as -669 mV in the pH range 4-10. Amperometric measurement could also be employed to monitor the pH of aqueous solution and the response sensitivity (dI/dpH) is similar to 67 nA. A two-signal (i.e., potential and current) synchronous detection technology has been exploited for pH monitoring. (C) 2011 Elsevier Ltd. All rights reserved.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available