4.6 Article

An electrochemically preanodized screen-printed carbon electrode for achieving direct electron transfer to glucose oxidase

Journal

ELECTROCHEMISTRY COMMUNICATIONS
Volume 10, Issue 7, Pages 1094-1097

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.elecom.2008.05.020

Keywords

glucose oxidase; direct electron transfer; preanodization; disposable; screen-printed carbon electrode

Ask authors/readers for more resources

Here we report the unique property of a preanodized screen-printed carbon electrode (SPCE*) that can allow direct electron transfer (DET) reaction of glucose oxidase (GOx). The GOx can be immobilized in the composite of oxygen functionalities and edge plane sites generated during preanodization without additional cross-linking agents. The electron transfer rate of GOx is greatly enhanced to 4.38 s(-1) as a result of the conformational change of GOx in the microenvironment enabling the accessibility of active site for GOx to the electrode. The analytical versatility is further improved with the aid of Nafion film. As a consequence, the as-prepared electrode can be used as a glucose biosensor and the number of potential foreign species is then restricted by molecular size, permeation and/or (bio)chemical reaction. Most importantly, the disposable nature of the proposed electrode is expected to promote the DET-related researches. (C) 2008 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available