Electrochemical lead hydride generation; Atomic absorption spectrometry; Lead; Urine samples
Categories
Funding
- Universidad Simon Bolivar (USB), Caracas, Venezuela
Ask authors/readers for more resources
An electrochemical lead volatile species generation system, as a means for sample introduction into a flame atomic absorption spectrometer, has been developed and evaluated for the determination of lead in urine samples. The reaction cell, designed and manufactured in our laboratory, consists of a reaction compartment housing a reticulated glassy carbon cathode and a platinum wire anode. The cell can easily be coupled to the spectrometer via a gas-liquid separator. The characteristic of the cathode material, the volatile species generation efficiency and the possible interferences of some concomitants have been studied. Calculated detection limit based on the variability of a blank solution (3 s(b) criterion) for ten measurements was 11 mu g L-1 and the sensitivity determined from the slope of the calibration graph, was 0.152 L mg(-1). The reproducibility (RSD) for ten replicate measurements at 1.0 mg L-1 lead level, was 1.4%. The accuracy of the method was determined through the analysis of spike urine samples. Recovery of 103.71% +/- 0.05 was achieved.