4.5 Article

Simultaneous determination of acetaminophen and ascorbic acid at an unmodified boron-doped diamond electrode by differential pulse voltammetry in buffered media

Journal

ELECTROANALYSIS
Volume 20, Issue 12, Pages 1346-1353

Publisher

WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.200804188

Keywords

ascorbic acid; acetaminophen; buffered media; differential pulse voltammetry; boron-doped diamond electrode

Ask authors/readers for more resources

A boron-doped diamond electrode (BDDE) was used for the simultaneous anodic determination Of L-ascorbic acid (AA) and acetaminophen (AC) in aqueous buffered media by differential pulse voltammetry (DPV). Linear calibration plots of anodic current peaks versus concentration were obtained for both analytes in the concentration range 0.01 - 0.1 mM with very high correlation coefficients. RSD of 2 - 3% and high sensitivities were obtained from DPV data in single and dicomponent systems. The potential applicability of the DPV technique associated with standard addition was illustrated by simultaneous determination of AA and AC in real sample solutions made up from pharmaceutical products.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available