Retention of [18F]fluoride on reversed phase HPLC columns

Journal

JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS
Volume 111, Issue -, Pages 209-214

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jpba.2015.04.009

Keywords

[F-18]Fluoride recovery; HPLC; PET; Quality control; Radiopharmaceuticals

Ask authors/readers for more resources

As [F-18]fluoride is a starting reagent in the radiosynthesis of most fluorine-18 labeled positron emission tomography (PET) tracers, its chromatographic behavior on reversed phase (RP) HPLC columns is important for the purification performance and accuracy of RP HPLC quality control methods. We have investigated the chromatographic behavior and recovery of [F-18]fluoride as a function of the type and brand of RP HPLC column, the pH and the composition of the mobile phase. Elution and elution profile of [F-18]fluoride from six RP-HPLC columns (Waters XBridge C18 3 mm x 100 mm 3.5 mu m; Grace Platinum EPS C18 4.6 mm x 100 mm, 3 mu m; Waters)(Terra C18 4.6 mm x 250 mm, 5 mu m; Phenomenex C18 4.6 mm x 150 mm, 5 mu m; Hamilton PRP-1 column 4.1 mm x 150 mm, 5 mu m; Merck KGaA Chromolith Performance C18 3 mm x 100 mm) eluted with mobile phase composed of phosphate or acetate buffers (pH 2, 3, 4, 5, 73 and 9) and acetonitrile or ethanol as organic modifier were characterized. The elution profile was determined by on-line radioactivity measurement in the column eluate and recovery was calculated by comparison of radioactivity eluted with the HPLC column present or absent in the chromatographic flow path. Interestingly, [F-18]fluoride recovery increased with increasing pH. At pH 3 all packed silica-based columns showed significant retention of fluorine-18, whereas almost no retention was observed on a polymeric PRP-I column. However at pH 5, [F-18]fluoride recovery was above 90% for each tested column. In addition, small differences were observed when changing the composition of the mobile phase. We therefore recommend to use a mobile phase with pH > 5 for silica based C18 columns for both quality control and semi-preparative HPLC of fluorine-18 labeled PET radiopharmaceuticals. If required a lower pH can be used in combination with a polymer based HPLC column. (C) 2015 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available