Real-time monitoring of non-viable airborne particles correlates with airborne colonies and represents an acceptable surrogate for daily assessment of cell-processing cleanroom performance

Journal

CYTOTHERAPY
Volume 14, Issue 9, Pages 1144-1150

Publisher

INFORMA HEALTHCARE
DOI: 10.3109/14653249.2012.698728

Keywords

airborne colonies; airborne particulates; cleanroom; current good manufacturing practices; environmental monitoring

Funding

  1. Production Assistance for Cellular Therapy (PACT) [N01-HB-37165]

Ask authors/readers for more resources

Background aims. Airborne particulate monitoring is mandated as a component of good manufacturing practice. We present a procedure developed to monitor and interpret airborne particulates in an International Organization for Standardization (ISO) class 7 cleanroom used for the cell processing of Section 351 and Section 361 products. Methods. We collected paired viable and non-viable airborne particle data over a period of 1 year in locations chosen to provide a range of air quality. We used receiver operator characteristic (ROC) analysis to determine empirically the relationship between non-viable and viable airborne particle counts. Results. Viable and non-viable particles were well-correlated (r(2) = 0.78), with outlier observations at the low end of the scale (non-viable particles without detectable airborne colonies). ROC analysis predicted viable counts >= 0.5/feet(3) (a limit set by the United States Pharmacopeia) at an action limit of >= 32 000 particles (>= 0.5 mu)/feet(3), with 95.6% sensitivity and 50% specificity. This limit was exceeded 2.6 times during 18 months of retrospective daily cleanroom data (an expected false alarm rate of 1.3 times/year). After implementing this action limit, we were alerted in real time to an air-handling failure undetected by our hospital facilities management. Conclusions. A rational action limit for non-viable particles was determined based on the correlation with airborne colonies. Reaching or exceeding the action limit of 32 000 non-viable particles/feet(3) triggers suspension of cleanroom cell-processing activities, deep cleaning, investigation of air handling, and a deviation management process. Our full procedure for particle monitoring is available as an online supplement.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available