Comparison of Human Tear Film Osmolarity Measured by Electrical Impedance and Freezing Point Depression Techniques

Journal

CORNEA
Volume 29, Issue 9, Pages 1036-1041

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/ICO.0b013e3181cd9a1d

Keywords

tear film; dry eye; osmolarity

Categories

Funding

  1. OcuSense, Inc.

Ask authors/readers for more resources

Purpose: Tear hyperosmolarity is diagnostic of dry eye disease (DED), yet difficulty in measurement has limited its utility; development of new instruments could facilitate its clinical application. This study compares the new OcuSense TearLab osmometer (OcuSense, Inc, San Diego, CA), based on electrical impedance lab-on-a-chip'' nanoliter technology, with the freezing point depression Clifton Osmometer (Clifton Technical Physics, Hartford, NY). Methods: Thirty-six subjects were recruited: 15 DED (9 women, 6 men age: 41 +/- 16 years) and 21 controls (12 women, 9 men age: 35 +/- 12 years); criteria for DED were noninvasive tear breakup time <10 seconds, Schirmer I test <5 mm, and positive symptoms. Samples were collected from the inferior tear meniscus for testing with both osmometers. Results: Osmolarity values measured with OcuSense TearLab were 308 +/- 6 and 321 +/- 16 mOsm/L for controls and dry eye, respectively, and those measured with Clifton were 310 +/- 7 and 323 +/- 14 mOsm/L for controls and dry eye, respectively; these values were significantly different. Significant correlation was found between OcuSense and Clifton measurements (r = 0.904; P = 0.006). Bland-Altman analysis revealed agreement between techniques; the majority of points fell within the 95% confidence limits, and actual values differed by less than 1%. A cutoff value of >316 mOsm/L, derived from the distribution of osmolarity values, was used to diagnose DED with an effectiveness of 73% sensitivity, 90% specificity, and 85% positive predictive value for the OcuSense and 73% sensitivity, 71% specificity, and 65% positive predictive value for the Clifton in the study samples. Conclusions: Tear film osmolarity measured with the OcuSense TearLab system correlates well with the Clifton Osmometer. The new instrument has the potential to provide clinicians with a readily available clinically applicable measure, which could become the gold standard in DED.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available