Background Reconstruction rings and bone allografts have been proposed to manage severe acetabular bone loss. However, a high early failure rate of the Graft Augmentation Prosthesis (GAP) II reinforcement ring (Stryker Orthopaedics, Mahwah, NJ, USA) has been reported in one small series. Questions/Purposes We therefore determined (1) the survival of this device in combination with impacted morselized allograft bone in patients with severe defects and (2) the complication rate. Methods We retrospectively reviewed 24 patients (21 aseptic and three septic) with severe acetabular bone loss (10 hips with Type III defects and 14 with Type IV defects according to the American Academy of Orthopaedic Surgeons classification). We determined function and numbers of failures. The minimum followup was 24 months (mean, 34 months; range, 24-72 months). Results At latest followup, the reconstruction had failed in nine of the 24 patients: six with aseptic loosening, three with infection. The average postoperative Merle d'Aubigne-Postel score of the patients whose reconstructions had not failed was 16.6 points; at latest followup, these patients had radiographic evidence of incorporation and consolidation of bone allografts. Seven of the nine patients whose reconstructions had failed underwent reoperation. Fatigue fracture of the ring at the plate-cup union occurred in five patients at an average of 45 months postoperatively. All patients with failed reconstructions who underwent reoperation were treated with Trabecular Metal TM (Zimmer Inc, Warsaw, IN, USA) cups and were functioning well at latest followup. Conclusions We observed a high rate (37%) of early catastrophic failures of the GAP II reconstruction ring, particularly in patients with Type IV defects. Due to this high failure rate, we have abandoned its use.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available