4.5 Article

Accuracy of a newly developed integrated system for dental implant planning

Journal

CLINICAL ORAL IMPLANTS RESEARCH
Volume 20, Issue 11, Pages 1191-1199

Publisher

WILEY-BLACKWELL PUBLISHING, INC
DOI: 10.1111/j.1600-0501.2009.01764.x

Keywords

computer-aided surgery (CAS); cone-beam imaging (CBCT); dental implantology; surgical templates; drill templates; surgical guides

Ask authors/readers for more resources

Objectives To evaluate the accuracy of the first integrated system for cone-beam CT (CBCT) imaging, dental implant planning and surgical template-aided implant placement. Materials and methods On the basis of CBCT scans, a total of 54 implant positions were planned for 10 partially edentulous anatomical patient-equivalent models. Surgical guides were ordered from the manufacturer (SICAT). Two different types of guidance were assessed: for assessment of the SICAT system inherent accuracy vendor's titanium sleeves of 2 mm internal diameter and 5 mm length were utilized for pilot drills. The guide sleeves of the NobelGuide system were implemented for fully guided surgery and implant insertion. Deviations perpendicular to the implant axes at the crestal and apical end, as well as the angle deviations between the virtual planning data and the surgical results, were measured utilizing a follow-up CBCT investigation and referential marker-based registration. Results The SICAT system inherent mean deviation rates for the drilled pilot osteotomies were determined to be smaller than 500 mu m even at the apical end. Mean angle deviations of 1.18 degrees were determined. Utilizing the NobelGuide sleeve-in-sleeve system for fully guided implant insertion in combination with the investigated template technology enabled to insert dental implants with the same accuracy. Crestal deviations, in general, were significantly lower than the apical deviations. Conclusion Although hardly comparable due to different study designs and measurement strategies, the investigated SICAT system's inherent accuracy corresponds to the most favourable results for computer-aided surgery systems published so far. In combination with the NobelGuide surgical set for fully guided insertion, the same accuracy level could be maintained for implant positioning. To cite this article:Dreiseidler T, Neugebauer J, Ritter L, Lingohr T, Rothamel D, Mischkowski RA, Zoller JE. Accuracy of a newly developed integrated system for dental implant planning.Clin. Oral Impl. Res. 20, 2009; 1191-1199.doi: 10.1111/j.1600-0501.2009.01764.x.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available