4.2 Article

Building and Evaluating the Autoverification of Coagulation Items in the Laboratory Information System

Journal

CLINICAL LABORATORY
Volume 60, Issue 1, Pages 143-150

Publisher

CLIN LAB PUBL
DOI: 10.7754/Clin.Lab.2013.130109

Keywords

laboratory automation; turnaround time; coagulation; delta check; Laboratory Information System

Ask authors/readers for more resources

Background: The increasing number of specimens, the decreasing number of proficient staff, and clinical demands make autoverification important in the laboratory's future development. However, although autoverification has been widely applied, a full coverage of its technique is hard to achieve in a laboratory. The establishment of the autoverification rules and parameters is still unclear. Therefore, the aim of this study was to fill the vacancy by the method described below. Methods: First, all logic processes and autoverification rules were established with reference to the Clinical and Laboratory Standards Institute (CLSI) document Auto10-A-Autoverification of Clinical Laboratory Test Results-Approved Guideline. Second, we established rules in the LIS, collected clinical specimens for validation, and assessed the results in a live environment. Results: The whole passing rate of autoverification was 82.0% and the manual intervention rate was 18.0%. The highest passing rate of a single project was activated partial thromboplastin time (APTT), approximately 96.1%, followed by prothrombin time (PT), approximately 95.1%, and fibrinogen, approximately 90.9%. The turnaround time (TAT) was shortened by 31.8% (90 minutes vs. 132 minutes). Conclusions: Through implementing the autoverification, which accelerated verification and decreased the TAT And the odds of human review errors in the released results, we can save more time and concentrate on verifying the abnormal results and proceeding emergency tests.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available