4.7 Article

Three of 7 Hemoglobin A1c Point-of-Care Instruments Do Not Meet Generally Accepted Analytical Performance Criteria

Journal

CLINICAL CHEMISTRY
Volume 60, Issue 8, Pages 1062-1072

Publisher

OXFORD UNIV PRESS INC
DOI: 10.1373/clinchem.2014.224311

Keywords

-

Funding

  1. Siemens Medical Solutions Diagnostics USA
  2. Menarini Diagnostics Italy
  3. Alere Technologies Norway
  4. DiaSys Germany
  5. Quotient U.K.
  6. Roche Diagnostics Switzerland

Ask authors/readers for more resources

BACKGROUND: In 2009, we investigated the conformance of 8 hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments. Since then, instruments have improved and new devices are available on the market. In this second study, we evaluated the performance of DCA Vantage, Afinion, InnovaStar, Quo-Lab, Quo-Test, Cobas B101, and B-analyst Hb A(1c) POC instruments. METHODS: Clinical and Laboratory Standards Institute protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using the mean of 3 certified secondary reference measurement procedures (SRMPs). Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria was also evaluated. Interference of common Hb variants was investigated for methods that could work with hemolysed material. RESULTS: The total CVs for all instruments, except for the DCA Vantage at a high Hb A(1c) value, were <= 3.1% in SI units and <= 2.1% in Diabetes Control and Complications Trial (DCCT) units. Afinion, DCA Vantage, B-analyst, and Cobas B101 instruments passed the NGSP criteria with 2 different reagent lot numbers. Quo-Test, Quo-Lab, and InnovaStar instruments had a negative bias compared to the mean of the 3 SRMPs and failed NGSP criteria. Most of the common Hb variants did not interfere with the investigated instruments, except Hb AE for the Cobas B101. CONCLUSIONS: Afinion, DCA Vantage, Cobas B101, and B-analyst instruments met the generally accepted performance criteria for Hb A(1c). Quo-Test, Quo-Lab, and InnovaStar met the criteria for precision but not for bias. Proficiency testing should be mandated for users of Hb A(1c) POC assays to ensure quality. (C) 2014 American Association for Clinical Chemistry

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available