Simple Paper-Based Test for Measuring Blood Hemoglobin Concentration in Resource-Limited Settings

Journal

CLINICAL CHEMISTRY
Volume 59, Issue 10, Pages 1506-1513

Publisher

AMER ASSOC CLINICAL CHEMISTRY
DOI: 10.1373/clinchem.2013.204701

Keywords

-

Funding

  1. NIH Director's Transformative Research Award [NHLBI R01HL117329]

Ask authors/readers for more resources

BACKGROUND: The measurement of hemoglobin concentration ([Hb]) is performed routinely as a part of a complete blood cell count to evaluate the oxygen-carrying capacity of blood. Devices currently available to physicians and clinical laboratories for measuring [Hb] are accurate, operate on small samples, and provide results rapidly, but may be prohibitively expensive for resource-limited settings. The unavailability of accurate but inexpensive diagnostic tools often precludes proper diagnosis of anemia in low-income developing countries. Therefore, we developed a simple paper-based assay for measuring [Hb]. METHODS: A 20-mu L droplet of a mixture of blood and Drabkin reagent was deposited onto patterned chromatography paper. The resulting blood stain was digitized with a portable scanner and analyzed. The mean color intensity of the blood stain was used to quantify [Hb]. We compared the performance of the paper-based Hb assay with a hematology analyzer (comparison method) using blood samples from 54 subjects. RESULTS: The values of [Hb] measured by the paper-based assay and the comparison method were highly correlated (R-2 = 0.9598); the standard deviation of the difference between the two measurements was 0.62 g/dL. The assay was accurate within 1 g/dL 90.7% of the time, overestimating [Hb] by >= 1 g/dL in 1.9% and underestimating [Hb] by >= 1 g/dL in 7.4% of the subjects. CONCLUSIONS: This study demonstrates the feasibility of the paper-based Hb assay. This simple, low-cost test should be useful for diagnosing anemia in resource-limited settings, particularly in the context of care for malaria, HIV, and sickle cell disease patients in sub-Saharan Africa. (c) 2013 American Association for Clinical Chemistry

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available