Performance evaluation of Siemens ADVIA Centaur® enhanced estradiol assay and a split sample comparison with liquid chromatography-tandem mass spectrometry

Journal

CLINICAL BIOCHEMISTRY
Volume 45, Issue 10-11, Pages 811-815

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2012.02.016

Keywords

Estradiol; Immunoassay; Method comparison; LC-MS/MS

Funding

  1. NCATS NIH HHS [UL1 TR000101] Funding Source: Medline

Ask authors/readers for more resources

Objectives: The objective of this study was to evaluate the newly-developed Siemens ADVIA Centaur (R) enhanced Estradiol (eE2) assay and compare it with a well-established estradiol liquid chromatography-tandem mass spectrometry (LC-MS/MS) method. Design and methods: The Siemens eE2 assay was evaluated using the Clinical and Laboratory Standards Institute evaluation protocols. Split patient samples were compared with the eE2 assay, the current ADVIA Centaur E2-6 III assay; and LC-MS/MS method by API5000 mass spectrometer. Results: Within-run and total imprecision of the eE2 assay demonstrated coefficient of variations of 5.7%, 3.2%, 1.5%, and 10.4%, 7.3%, and 6.8%, at levels of 380, 752, and 2051 pmol/L, respectively. The method comparisons showed: eE2=0.903(E2-6 III) -16.2, R-2=0.938, average bias= -12.3%; and eE2=0.946(LC-MS/MS) + 19.5, R-2=0.925, average bias: 0%. Conclusion: The Siemens eE2 assay correlates well with LC-MS/MS. This method is reliable, and appropriate for routine clinical laboratory use. (C) 2012 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available