3.9 Article

Good Performance of Immunoglobulin M Assays in Diagnosing Genotype 3 Hepatitis E Virus Infections

Journal

CLINICAL AND VACCINE IMMUNOLOGY
Volume 16, Issue 5, Pages 772-774

Publisher

AMER SOC MICROBIOLOGY
DOI: 10.1128/CVI.00438-08

Keywords

-

Ask authors/readers for more resources

We have evaluated three anti-hepatitis E virus (anti-HEV) immunoglobulin M (IgM) assays, the EIAgen HEV IgM assay (Adaltis), the HEV IgM enzyme-linked immunosorbent assay 3.0, and the Assure HEV IgM rapid test (MP Diagnostics), for the routine detection of acute genotype 3 HEV. Their sensitivities were fairly good (90%, 88%, and 82%), and their specificities were excellent (100%, 99.5%, and 100%).