4.4 Article Proceedings Paper

Accurate calibration of test mass displacement in the LIGO interferometers

Journal

CLASSICAL AND QUANTUM GRAVITY
Volume 27, Issue 8, Pages -

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/0264-9381/27/8/084024

Keywords

-

Funding

  1. Direct For Mathematical & Physical Scien
  2. Division Of Physics [0757058] Funding Source: National Science Foundation

Ask authors/readers for more resources

We describe three fundamentally different methods we have applied to calibrate the test mass displacement actuators to search for systematic errors in the calibration of the LIGO gravitational-wave detectors. The actuation frequencies tested range from 90 Hz to 1 kHz and the actuation amplitudes range from 10(-6) m to 10(-18) m. For each of the four test mass actuators measured, the weighted mean coefficient over all frequencies for each technique deviates from the average actuation coefficient for all three techniques by less than 4%. This result indicates that systematic errors in the calibration of the responses of the LIGO detectors to differential length variations are within the stated uncertainties.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available