4.3 Article

Long-Term Safety of an Everolimus-Eluting Bioresorbable Vascular Scaffold and the Cobalt-Chromium XIENCE V Stent in a Porcine Coronary Artery Model

Journal

CIRCULATION-CARDIOVASCULAR INTERVENTIONS
Volume 7, Issue 3, Pages 330-342

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1161/CIRCINTERVENTIONS.113.000990

Keywords

coronary disease; pathology; stents

Funding

  1. Abbott Vascular
  2. CVPath Institute, Inc
  3. Uehara Memorial Foundation, Tokyo, Japan

Ask authors/readers for more resources

Background The Absorb everolimus-eluting bioresorbable vascular scaffold (Absorb) has shown promising clinical results; however, only limited preclinical data have been published. We sought to investigate detailed pathological responses to the Absorb versus XIENCE V (XV) in a porcine coronary model with duration of implant extending from 1 to 42 months. Methods and Results A total of 335 devices (263 Absorb and 72 XV) were implanted in 2 or 3 main coronary arteries of 136 nonatherosclerotic swine and examined by light microscopy, scanning electron microscopy, pharmacokinetics, and gel permeation chromatography analyses at various time points. Vascular responses to Absorb and XV were largely comparable at all time points, with struts being sequestered within the neointima. Inflammation was mild to moderate (with absence of inflammation at 1 month) for both devices, although the scores were greater in Absorb at 6 to 36 months. Percent area stenosis was significantly greater in Absorb than XV at all time points except at 3 months. The extent of fibrin deposition was similar between Absorb and XV, which peaked at 1 month and decreased rapidly thereafter. Histomorphometry showed expansile remodeling of Absorb-implanted arteries starting after 12 months, and lumen area was significantly greater in Absorb than XV at 36 and 42 months. These changes correlated with dismantling of Absorb seen after 12 months. Gel permeation chromatography analysis confirmed that degradation of Absorb was complete by 36 months. Conclusions Absorb demonstrates comparable long-term safety to XV in porcine coronary arteries with mild to moderate inflammation. Although Absorb was associated with greater percent stenosis relative to XV, expansile remodeling was observed after 12 months in Absorb with significantly greater lumen area at 36 months. Resorption is considered complete at 36 months.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available