4.8 Article

Accelerated Degeneration of a Bovine Pericardial Bioprosthetic Aortic Valve in Children and Young Adults

Journal

CIRCULATION
Volume 130, Issue 1, Pages 51-+

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1161/CIRCULATIONAHA.114.009835

Keywords

heart defects, congenital; prosthesis implantation; valves

Funding

  1. Farb Family Fund

Ask authors/readers for more resources

Background-Experience with aortic valve replacement (AVR) with current-generation pericardial bioprostheses in young patients is limited. The death of a child with accelerated bioprosthetic aortic stenosis prompted enhanced surveillance of all such patients at our institution. Methods and Results-We reviewed records of 27 patients who had undergone AVR (median follow-up, 13.7 months) with a bovine pericardial bioprosthesis at <= 30 years of age. In the Mitroflow LXA valve group (n=15), freedom from valve failure was 100% at 1 year, 53% (95% confidence interval, 12-82) at 2 years, and 18% (95% confidence interval, 1-53) at 3 years. No Magna/Magna Ease valves (n=12) failed by 3 years. Among valve failure patients, median age at AVR was 12 years (range, 10-21 years). Life-threatening prosthetic aortic stenosis was detected at a median of 6 months after prior echocardiograms showing mild or less gradients. Patients with Mitroflow LXA compared with Magna/Magna Ease valves were smaller (median body surface area, 1.42 versus 1.93 m(2); P=0.002) and younger (median age, 13.0 versus 20.9 years; P=0.02) at AVR. Pathology demonstrated diffuse intrinsic leaflet calcification, not associated with inflammation or infection, and virtually immobile leaflets in closed position. Conclusions-Young patients undergoing AVR with Mitroflow LXA pericardial valves are at high risk for rapid progression from mild or less to severe aortic stenosis over months, highlighting their need for heightened echocardiographic surveillance and suggesting that this aortic bioprosthesis should not be implanted in the young. Current data are insufficient to assess the safety of AVR with other pericardial bioprostheses in children and the youngest adults.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available