4.5 Article

Mechanical complications of dental implants

Journal

CLINICAL ORAL IMPLANTS RESEARCH
Volume 11, Issue -, Pages 156-158

Publisher

MUNKSGAARD INT PUBL LTD
DOI: 10.1034/j.1600-0501.2000.011S1156.x

Keywords

dental implants; mechanical; complications; failures; interfaces; screw loosening; fracture

Ask authors/readers for more resources

Adverse occlusal forces can result in mechanical complications of implant components. While unacceptably high incidences of mechanical failures have been reported for the two-stage external hex screw-type implant systems, the standard-diameter ITI(R) solid-screw implant does not seem to be vulnerable to these problems. The 8 degrees Morse taper has eliminated abutment screw loosening and fracture. The incidence of prosthetic screw loosening has been minimized by the 45 degrees bevel on the implant shoulder and by the 1.5 mm vertical abutment walls. The design of the standard-diameter solid-screw ITI implant and the material used in its fabrication (cold worked type IV cp titanium) have eliminated fixture fracture. However, because there have been some reported instances of fractures involving reduced-diameter and hollow implants, these designs should be used with caution.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available