A novel method for the prediction of functional biological activity of polyethylene wear debris

Publisher

SAGE PUBLICATIONS LTD
DOI: 10.1243/0954411011533599

Keywords

joint replacement; polyethylene; wear debris; functional biological activity; model

Ask authors/readers for more resources

The comparative performance of artificial hip joints has been extensively investigated in vitro through measurements of wear volumes. In vivo a major cause of long-term failure is wear-debris-induced osteolysis. These adverse biological reactions are not simply dependent on wear volume, but are also controlled by the size and volumetric concentration of the debris. A novel model is presented which predicts functional biological activity; this is determined by integrating the product of the biological activity function and the volumetric concentration function with the wear volume over the whole particle size range. This model combines conventional wear volume measurements with particle analysis and the output from in vitro, cell culture studies to provide a new indicator of osteolytic potential. The application of the model is demonstrated through comparison of the functional biological activity of wear debris from polyethylene acetabular cups articulating under three different conditions in a hip joint simulator.