Journal
ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 372, Issue 2, Pages 284-292Publisher
SPRINGER-VERLAG BERLIN
DOI: 10.1007/s00216-001-1128-3
Keywords
amperometry; creatinine; creatine; hemodialysis; screen-printed electrode; biosensor
Ask authors/readers for more resources
Biosensors for the determination of creatinine have been developed and integrated into a flow-through system. The sensors are based on a screen-printed three electrode transducer with a platinum working electrode. Applying the multi-enzyme sequence of creatininase (CA), creatinase (CI) and sarcosine oxidase (SO) hydrogen peroxide has been detected amperometrically. An optimal enzyme load was found to be 4.4 U/0.28 U/0.20 U (CA/CI/SO) and 0.28 U/0.20 U (CI/SO) per electrode for the creatinine sensor and for the creatine sensor, respectively. Among a variety of polymers Nafion has shown the highest efficiency to exclude interfering substances like ascorbic acid, acetaminophen and uric acid. First determinations of creatinine in dialysate samples obtained during hemodialysis treatments have shown a good correlation to the conventional methods, the Jaffe reaction (y=0.945x+ 2.8, R=0.9882, n=9) and the enzymatic photometric method (y=0.891x+3.5, R=0.9917, n=9).
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available