4.4 Article

Determination of monocrotophos pesticide by flow injection chemiluminescence method using luminol-hydrogen peroxide system

Journal

ANALYTICAL LETTERS
Volume 36, Issue 5, Pages 1029-1038

Publisher

TAYLOR & FRANCIS INC
DOI: 10.1081/AL-120019260

Keywords

chemiluminescence; flow injection; monocrotophos pesticide

Ask authors/readers for more resources

A simple flow injection chemiluminescence method is described for the determination of monocrotophos pesticide, based upon the direct reaction of monocrotophos with luminol and hydrogen peroxide in alkaline medium in the presence of an enhancer of sodium chloride. The CL intensity was linear related to monocrotophos concentration in the range of 2.0 x 10(-8)-1.0 x 10(-6) g/mL with a detection limit of 7 x 10(-9) g/mL. The relative standard deviation was less than 3% for 1.0 x 10(-7) g/mL monocrotophos in eleven repeated measurements. This method was successfully applied to the determination of monocrotophos in water samples. The possible reaction mechanism was also discussed briefly.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available