4.6 Article

The determination of trace metals in estuarine and coastal waters using a voltammetric in situ profiling system

Journal

ANALYST
Volume 128, Issue 6, Pages 734-741

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/b300712j

Keywords

-

Ask authors/readers for more resources

This work presents the optimisation, validation and field deployment of a voltammetric in situ profiling (VIP) system for the simultaneous determinations of dynamic Cd(II), Cu(II) and Pb(II) in estuarine and coastal waters. Systematic studies in NaNO3 (as a supporting electrolyte) and seawater, indicated that variations in ionic strength, pH and dissolved oxygen did not affect the response of the instrument, whereas an Arrhenius type temperature response was observed. The VIP instrument allows the determination of 2-3 samples h(-1), and has a detection limit (defined as 3sigma) in seawater for Cd(II): 23 pM, Cu(II): 1.13 nM, and Pb(II): 23 pM. The VIP system accurately measured the total dissolved concentrations of Cd(II), Cu(II) and Pb(II) in two certified reference materials; SLRS-3, a river water, and SLEW-2, an estuarine water. Field evaluation of the instrumentation and analytical methods was achieved through a series of surveys in the Plym Estuary (Devon, UK), from which environmental data are presented.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available