4.6 Article

Advanced optical tweezers for the study of cellular and molecular biomechanics

Journal

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
Volume 50, Issue 1, Pages 121-125

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2002.805463

Keywords

biomechanics; calibration; optical tweezers

Ask authors/readers for more resources

Optical tweezers are an important tool for studying cellular and molecular biomechanics. We present a robust optical tweezers device with advanced features including: multiple optical traps, acousto-optic trap steering, and back focal plane interferometry position detection. We integrate these features into an upright microscope, with no compromise to its capabilities (differential interference contrast microscopy, fluorescence microscopy, etc.). Acousto-optic deflectors (AODs) steer each beam and can create multiple time-shared traps. Position detection, force calibrations and AOD performance are presented. The system can detect subnanometer displacements and forces below 0.1 pN.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available