4.6 Article

A fast method for the measurement of long spin-lattice relaxation times by single scan inversion recovery experiment

Journal

CHEMICAL PHYSICS LETTERS
Volume 383, Issue 1-2, Pages 99-103

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.cplett.2003.10.133

Keywords

-

Ask authors/readers for more resources

A new method of measuring long spin-lattice relaxation times (T-1) is proposed. Being a single scan technique, the method is at least one order of magnitude faster than the conventional technique. This method (single-scan or slice selected inversion recovery or SSIR) relies on the slice selection technique. The method is experimentally verified and compared with the time taken by the conventional measurement. Furthermore, it is shown that the conventional Inversion Recovery method which suffers from effects of r.f. inhomogeneity can also be improved by measuring the magnetization of only a central slice. (C) 2003 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available