4.3 Article

A continuous-wave and pulsed electron spin resonance spectrometer operating at 275 GHz

Journal

JOURNAL OF MAGNETIC RESONANCE
Volume 166, Issue 1, Pages 92-99

Publisher

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jmr.2003.10.011

Keywords

EPR; ESE; high field; high-frequency; mm-wave

Ask authors/readers for more resources

An electron paramagnetic resonance (EPR) spectrometer is described which allows for continuous-wave and pulsed EPR experiments at 275 GHz (wavelength 1.1 mm). The related magnetic field of 9.9 T for g similar to 2 is supplied by a superconducting solenoid. The microwave bridge employs quasi-optical as well as conventional waveguide components. A cylindrical, single-mode cavity provides a high filling factor and a high sensitivity for EPR detection. Even with the available microwave power of 1 mW incident at the cavity a high microwave magnetic field B-1 is obtained of about 0.1 mT which permits pi/2-pulses as short as 100 ns. The performance of the spectrometer is illustrated with the help of spectra taken with several samples. (C) 2003 Published by Elsevier Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available