4.7 Article

Air driven on-line separation and preconcentration on a C-18 column coupled with thermospray flame furnace AAS for the determination of cadmium and lead at mu g l(-1) levels

Journal

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
Volume 19, Issue 11, Pages 1507-1509

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/b409182e

Keywords

-

Ask authors/readers for more resources

A simple and sensitive method for the determination of cadmium and lead at mug l(-1) and sub-mug l(-1) levels by on- line separation and preconcentration coupled with thermospray flame furnace AAS ( TS- FF- AAS) was developed. Compressed air was used to drive the sample through the 3 cm HPLC C-18 RP column and the 500 mul eluate to the TS- FF- AAS. The chelating reagent ammonium pyrrolidinedithiocarbamate was immobilized onto the column as a solution in the eluent ( 60% v/ v methanol) simultaneously with the elution of the previous preconcentrated sample. Loading of sample volumes up to 20 ml was possible. The detection limits for cadmium and lead using 20 ml sample volume were 0.007 and 0.17 mug l(-1), respectively. RSD values were typically about 4%. The combined effect of preconcentration and TS- FF- AAS yielded enhancement factors towards conventional flame AAS of 1960 and 610 for cadmium and lead, respectively.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available