4.6 Article

Determination of endocrine disruptors in environmental waters using poly(acrylamide-vinylpyridine) monolithic capillary for in-tube solid-phase microextraction coupled to high-performance liquid chromatography with fluorescence detection

Journal

ANALYST
Volume 130, Issue 7, Pages 1065-1069

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/b502311d

Keywords

-

Ask authors/readers for more resources

A method for the determination of endocrine disruptors, bisphenol A and 17 alpha-ethinylestradiol, in environmental water samples was developed using in-tube solid-phase microextraction followed by liquid chromatography and fluorescence detection. A poly(acrylamide-vinylpyridine) monolithic capillary column was applied as the extraction media in view of its greater phase ratio than open-tubular capillaries and thus higher extraction efficiency. After optimizing the extraction conditions, bisphenol A and 17 alpha-ethinylestradiol were extracted directly from water samples in a wide dynamic linear range of 0.5-1000 ng mL(-1), with the detection limits obtained as 0.064 and 0.12 ng mL(-1), respectively. The precision of the method was satisfactory with the intraday and interday RSD values smaller than 7.2%. Environmental water samples of different sources were successfully analyzed with the presented method and the monolithic capillary was proved to be robust and reusable in analyzing real water samples.