Journal
ANALYTICAL METHODS
Volume 7, Issue 16, Pages 6732-6739Publisher
ROYAL SOC CHEMISTRY
DOI: 10.1039/c5ay00801h
Keywords
-
Funding
- National Water Pollution Control and Management Technology Major Projects [2012ZX07307-001]
Ask authors/readers for more resources
A dual-cloud point extraction (d-CPE) technique combined with inductively coupled plasma mass spectrometry (ICP-MS) was used for preconcentration and determination of Cr, Ga, Ag, Cd, Mn, Fe, In, Cu, Ni, Co, Pb and Zn in real water samples. An orthogonal array design (OAD) with the OA(25) (5(6)) matrix was employed to study the effects of different parameters on the extraction efficiency. Under the best experimental conditions (concentration of complexing reagent 8-hydroxyquinoline, 0.25 mmol L-1; pH, 6.5; equilibrium temperature, 60 degrees C; surfactant Triton X-114, 0.1%; equilibrium time, 30 min and without salt addition), the enhancement factor ranged from 9.85 to 35.98. The linear dynamic range (LDR) was 1-800 mg L-1 with correlation coefficients of 0.9974-0.9999 and the limits of detection (LODs) were between 0.012 and 0.36 mg L-1. Relative standard deviations (RSDs, C = 10 mg L-1 and n = 6) were 1.29%-5.56%. Finally, the developed method was successfully applied to extraction and determination of the mentioned metal ions in real water samples and satisfactory results were obtained.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available