3.8 Article

Calibration of the spring constant of AFM cantilever

Journal

JOURNAL OF ELECTRON MICROSCOPY
Volume 56, Issue 1, Pages 21-25

Publisher

OXFORD UNIV PRESS
DOI: 10.1093/jmicro/dfm001

Keywords

spring constant; resonance frequency; cantilever; calibration; Q curve; atomic force microscopy

Categories

Ask authors/readers for more resources

Several methods have been developed in recent years to calibrate the spring constant of atomic force microscopy (AFM) cantilever. In the present work, an effective and practical improvement is made on these methods to provide a more convenient calibration method. Simplifying the originally tedious procedures by just examining the cantilever's dimensions in scanning electron microscopy (SEM) and the resonance frequency of the cantilever from Q curves in AFM system, the spring constants of different cantilevers are calculated by the methods of Cleveland and Sader. The results are close to the nominal values provided by the manufacturers, which indicates that simplifying the procedures is valid and reliable. Moreover, it shows that spring constants acquiring by Cleveland method is more close to their nominal values than Sader's method does.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

3.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available