4.5 Article

Direct measurement of cantilever spring constants and correction for cantilever irregularities using an instrumented indenter

Journal

REVIEW OF SCIENTIFIC INSTRUMENTS
Volume 78, Issue 6, Pages -

Publisher

AMER INST PHYSICS
DOI: 10.1063/1.2747095

Keywords

-

Ask authors/readers for more resources

A method is presented that allows direct measurement of a wide range of spring constants of cantilevers using an indentation instrument with an integrated optical microscope. An uncertainty of less than 10% can be achieved for spring constants from 0.1 to 10(2) N/m. The technique makes it possible to measure the spring constant at any desired location on a cantilever of any shape, particularly at the tip location of an atomic force microscopy cantilever. The article also demonstrates a technique to detect and correct apparent length anomalies of cantilevers by analyzing spring constants at multiple positions. (c) 2007 American Institute of Physics.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available