The application of ultra-performance liquid chromatography/tandem mass spectrometry to the detection and quantitation of apolipoproteins in human serum

Journal

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Volume 21, Issue 16, Pages 2585-2593

Publisher

JOHN WILEY & SONS LTD
DOI: 10.1002/rcm.3130

Keywords

-

Ask authors/readers for more resources

The detection and quantitation of apolipoproteins, important markers for coronary heart disease, in serum by liquid chromatography/tandem mass spectrometry (LC/MS/MS) using multiple reaction monitoring (MRM) is reported. A tryptic digest of depleted human serum was analysed by nanoflow LC/MS/MS at a flow rate of 300 nL/min and several apolipoproteins (Apo), including Apo A1, A2, A4, C1, C2, C3, D, F and M, were successfully identified. The analysis of the same depleted serum digest by ultra-performance (UP)LC/MS/MS operating at 700 mu L/min resulted in comparable sensitivity and selectivity to the nanoflow method, but with a dramatic (similar to 20-fold) reduction in run time. The potential of UPLC/MS/MS for the rapid quantitation of proteins in biological matrices by representative tryptic peptides was further investigated using Apo Al and its corresponding stable isotopically labelled tryptic AQUA peptide (DYVSQFEGSALGK). A set of serum-based Apo Al calibrators from a clinical analyser kit were digested without depletion following the addition of the AQUA peptide and analysed using UPLC/MS/MS. A linear calibration curve was generated from peak area ratios to the labelled peptide with a coefficient of correlation of 0.9989. Standard curves were also generated for other apolipoproteins together with Apo B100, Apo E, lecithin cholesterol acyltransferase and albumin, which were also detected in the standards. The concentration of Apo Al in five fresh undepleted human serum samples and a quality control (QC) sample were determined using both the UPLC/MS/MS method and a clinical analyser. Results were comparable and the quantitative study, involving 80 injections which took hours rather than days to complete, demonstrates the high-throughput potential of UPLC/MS/MS to quantify multiple serum proteins without the need for antibodies, and thus provide an alternative to the use of clinical analysers for serum protein biomarkers. Copyright (c) 2007 John Wiley & Sons, Ltd.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available