Glucose is a common medical analyte measuring in human serum or blood samples. The development of a primary method is necessary for the establishment of traceability in measurements. We have developed an isotope dilution liquid chromatography tandem mass spectrometry as a primary method for the measurement of glucose in human serum. Glucose and glucose-C-13(6) in sample were ionized in ESI negative mode and monitored at mass transfers of m/z 179/89 and 185/92 in MRM, respectively. Glucose was separated on NH2P-50 2D column, and the mobile phase was 20 mM NH4OAc in 30% acetonitrile/70% water. Verification of this method was performed by the comparison with NIST SRMs. Our results agreed well with the SRM values. We have developed two levels of glucose strum certified reference material using this method and distributed them to the clinical laboratories in Korea as samples for proficiency testings. The expended uncertainty was about 1.2% on 95% confidence level. In proficiency testings, the results obtained from the clinical laboratories showed about 3.6% and 3.9% RSD to the certified values. Primary method can provide the traceability to the field laboratories through proficiency testings or certified reference materials.