4.4 Article

Determination of Ractopamine and Clenbuterol in Beef by Graphene Oxide Hollow Fiber Solid-Phase Microextraction and High-Performance Liquid Chromatography

Journal

ANALYTICAL LETTERS
Volume 49, Issue 8, Pages 1163-1175

Publisher

TAYLOR & FRANCIS INC
DOI: 10.1080/00032719.2015.1094662

Keywords

Clenbuterol; graphene oxide; hollow fiber; ractopamine; solid-phase microextraction

Funding

  1. Importation and Development of High-Caliber Talents Project of Beijing Municipal Institutions [CIT TCD 201504083]
  2. National Natural Science Foundation of China [21005050]
  3. Beijing Natural Science Foundation [2162008]

Ask authors/readers for more resources

A novel solid-phase microextraction method is reported that involved modification of the pores of hollow fibers with graphene oxide. The modified graphene oxide/hollow fiber was characterized by scanning electron microscopy and Raman spectroscopy and employed for the determination of ractopamine and clenbuterol in fortified beef. The repeatability of the method was satisfactory with intra-day and inter-day relative standard deviations between 2.5 and 5.2% for ractopamine and clenbuterol. Under the optimized conditions, the linear dynamic ranges of ractopamine and clenbuterol were from 0.05 to 2.5 mu g/mL and 0.1 to 2.5 mu g/mL, respectively. The limits of detection for ractopamine and clenbuterol were 0.01 and 0.03 mu g/mL, and the limits of quantification were 0.03 and 0.08 mu g/mL. The developed method was used for the determination of ractopamine and clenbuterol in fortified beef with average recoveries of 85.8-109.8% and 78.8-101.4%.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available