4.1 Article

Comparison of microbiological assay and HPLC-UV for determination of fluconazole in capsules

Journal

BRAZILIAN JOURNAL OF PHARMACEUTICAL SCIENCES
Volume 45, Issue 4, Pages 693-700

Publisher

UNIV SAO PAULO, CONJUNTO QUIMICAS
DOI: 10.1590/S1984-82502009000400012

Keywords

Fluconazole/quantitative determination; Microbiological assay/agar diffusion; High performance liquid chromatography

Ask authors/readers for more resources

The development of a specific agar diffusion bioassay for the quantitative determination of fluconazole formulated in capsules was carried out using a strain of Candida albicans ATCC 18804 as the test organism. A prospective validation of the method showed adequate linearity (r(2)=0.9995), precision (R.S.D. = 4.0% for intra-day and 4.5% for inter-day precision) and accuracy (mean recovery = 102.9%). High performance liquid chromatography was chosen as a comparison method for the fluconazole determination. The contents of fluconazole determined by both methods, for four capsule samples, showed a strong correlation, confirmed by Pearson's correlation coefficient value (r = 0.9884). The bioassay is a suitable method for both research and pharmaceutical industry laboratories.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available