Validation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010

Journal

VASCULAR HEALTH AND RISK MANAGEMENT
Volume 9, Issue -, Pages 265-272

Publisher

DOVE MEDICAL PRESS LTD
DOI: 10.2147/VHRM.S44569

Keywords

blood pressure; device; European Society of Hypertension; guideline; measurement; validation

Ask authors/readers for more resources

Background: Allowing patients to measure their blood pressure at home is recognized as being of clinical value. However, it is not known how often these measurements are taken correctly. Blood pressure monitors for home use fall into two types based on the position of the cuff, ie, at the upper arm or the wrist. The latter is particularly convenient, as measurements can be taken fully clothed. This study aimed to evaluate the performance of the wrist-type blood pressure monitors Omron RS8 (HEM-6310F-E), Omron RS6 (HEM-6221-E), and Omron RS3 (HEM-6130-E). Methods: A team of three trained doctors validated the performance of these devices by comparing the measurements obtained from these devices with those taken using a standard mercury sphygmomanometer. All the devices met the validation requirements of the European Society of Hypertension International Protocol revision 2010. Results: The difference in blood pressure readings between the tested device and the standard mercury sphygmomanometer was within 3mmHg, which is acceptable according to the European Society of Hypertension guidelines. Conclusion: All the home devices tested were found to be suitable for measuring blood pressure at home because their performance fulfilled the requirement of the guidelines.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

3.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available