Journal
ANALYTICAL BIOCHEMISTRY
Volume 489, Issue -, Pages 9-16Publisher
ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.ab.2015.08.007
Keywords
Poly(glutamic acid)-modified electrode; Glucose oxidase; Interfacial electron transfer; Glucose sensing
Ask authors/readers for more resources
The interfacial electron transfer of glucose oxidase (GOx) on a poly(glutamic acid)-modified glassy carbon electrode (PGA/GCE) was investigated. The redox peaks measured for GOx and flavin adenine dinucleotide (FAD) are similar, and the anodic peak of GOx does not increase in the presence of glucose in a mediator-free solution. These indicate that the electroactivity of GOx is not the direct electron transfer (DET) between GOx and PGA/GCE and that the observed electroactivity of GOx is ascribed to free FAD that is released from GOx. However, efficient electron transfer occurred if an appropriate mediator was placed in solution, suggesting that GOx is active. The PGA/GCE-based biosensor showed wide linear response in the range of 0.5-5.5 mM with a low detection limit of 0.12 mM and high sensitivity and selectivity for measuring glucose. (C) 2015 Elsevier Inc. All rights reserved.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available