4.2 Article

Dental Strain on Maxillary Incisors During Tracheal Intubation With Double-Lumen Tubes and Different Laryngoscopy Techniques - A Blinded Mannequin Study

Journal

JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA
Volume 36, Issue 8, Pages 3021-3027

Publisher

W B SAUNDERS CO-ELSEVIER INC
DOI: 10.1053/j.jvca.2022.02.017

Keywords

anesthetic complications; dental trauma; difficult airway; double-lumen tube intubation; videolaryngoscopy

Funding

  1. University Witten/Herdecke Research support [IFF2017-09]

Ask authors/readers for more resources

The study quantified dental forces during double-lumen tube intubations with different laryngoscopy techniques. The results showed that while videolaryngoscopes reduced dental strain, they also prolonged the time to intubation. Different laryngoscopy techniques performed differently under normal and difficult airway conditions.
Objectives: To quantify dental forces during double-lumen tube intubations with different laryngoscopy techniques. Design: Experimental biomechanical mannequin study. Setting: Two German university hospitals. Participants: One hundred four anesthesiologists with varying levels of experience. Interventions: Participants performed a sequence of intubations on a mannequin equipped with hidden forces sensors in the maxillary incisors. Different laryngoscopy techniques were evaluated under normal and difficult airway conditions. Direct laryngoscopy was compared with different videolaryngoscopy techniques: the C-MAC with a Macintosh blade, the GlideScope, and the KingVision with hyperangulated blades. Measurements and Main Results: A total of 624 intubations were evaluated. In normal airway conditions, the median (interquartile range [range]) peak forces were significantly lower when the GlideScope (15.7 (11.3-22.0 [2.1-110.5]) N) was used compared with direct laryngoscopy (21.0 (14.1-28.5[4.7-168.6]) N) (p = 0.007). In difficult airways, resulting forces were reduced using hyperangulated videolaryngoscopes (Glide-Scope: -13.7 N [p < 0.001]; KingVision: -11.9 N [p < 0.001]) compared with direct laryngoscopy, respectively. The time to intubation was prolonged with the use of the KingVision (25.5 (17.1-41.9[9.2-275.0])s [p < 0.001]) in comparison to direct laryngoscopy (20.8 (15.9-27.4[8.7-198.6]) s). The C-MAC demonstrated the shortest time to intubation. Conclusions: Although hyperangulated videolaryngoscopes improve dental strain, clinicians also should consider the time to intubation, which is shortest with nonhyperangulated videoblades, when choosing a laryngoscopy technique on an individual patient basis. (C) 2022 Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available