4.3 Article

An electrically active microneedle array for electroporation

Journal

BIOMEDICAL MICRODEVICES
Volume 12, Issue 2, Pages 263-273

Publisher

SPRINGER
DOI: 10.1007/s10544-009-9381-x

Keywords

Microneedle; Electroporation; Micromolding; Laser ablation; DU145 cell

Funding

  1. National Institute of Health

Ask authors/readers for more resources

We have designed and fabricated a microneedle array with electrical functionality with the final goal of electroporating skin's epidermal cells to increase their transfection by DNA vaccines. The microneedle array was made of polymethylmethacrylate (PMMA) by micromolding technology from a polydimethylsiloxane (PDMS) mold, followed by metal deposition, patterning using laser ablation, and electrodeposition. This microneedle array possessed sufficient mechanical strength to penetrate human skin in vivo and was also able to electroporate both red blood cells and human prostate cancer cells as an in vitro model to demonstrate cell membrane permeabilization. A computational model to predict the effective volume for electroporation with respect to applied voltages was constructed from finite element simulation. This study demonstrates the mechanical and electrical functionalities of the first MEMS-fabricated microneedle array for electroporation, designed for DNA vaccine delivery.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available