4.4 Article

Development of a validated high-throughput LC-ESI-MS method for determination of sirolimus on dried blood spots

Journal

BIOMEDICAL CHROMATOGRAPHY
Volume 24, Issue 12, Pages 1356-1364

Publisher

JOHN WILEY & SONS LTD
DOI: 10.1002/bmc.1450

Keywords

sirolimus; dried blood spots; monolithic column; high-throughput; LC-ESI-MS

Funding

  1. University Grants Commission, New Delhi, India

Ask authors/readers for more resources

A high-throughput liquid chromatography-electrospray ionization mass spectrometric (LC-ESI-MS) method for screening of sirolimus on dried blood spots (DBS) was developed and validated. It involves solvent extraction of a punch of DBS followed by reversed-phase LC on a relatively new monolithic column consisting of a silica rod with bimodal pore structure and detection by ESI-MS. The run time was less than 3 min with a very low backpressure at a flow rate of 0.5 mL/min. The method can analyze more than 100 samples in an 8 h working day, including sample preparation. The assay was linear from 1 to 100 ng/mL. The mean recovery was 92.42%. The mean inter-day and intra-day precisions were 1.23 and 1.41%, respectively. The developed method is simple, rapid and useful for clinical applications. Copyright (C) 2010 John Wiley & Sons, Ltd.