Journal
BIOMEDICAL CHROMATOGRAPHY
Volume 24, Issue 7, Pages 683-691Publisher
WILEY
DOI: 10.1002/bmc.1346
Keywords
LC/MS; amino acids; derivatization; automated
Ask authors/readers for more resources
An automated method for high-throughput amino acid analysis, using precolumn derivatization high-performance liquid chromatography/electrospray mass spectrometry (HPLC/ESI-MS), was developed and evaluated. The precolumn derivatization step was performed in the reaction port of a home-built auto-sampler system. Amino acids were derivatized with 3-aminopyridyl-N-hydroxysuccinimidyl carbamate, and a 3 mu m Wakosil-II 3C8-100HG column (100 x 2.1 mm i.d.) was used for separation. To achieve a 13 min cycle for each sample, the derivatization and separation steps were performed in parallel. The results of the method evaluation, including the linearity, and the intra- and inter-precision, were sufficient to measure physiological amino acids in human plasma samples. The relative standard deviations of typical amino acids in actual human plasma samples were below 10%. Copyright (C) 2009 John Wiley & Sons, Ltd.