4.7 Article

Electrochemical detection of thrombin by sandwich approach using antibody and aptamer

Journal

BIOELECTROCHEMISTRY
Volume 73, Issue 1, Pages 76-81

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/j.bioelechem.2008.04.024

Keywords

protein; electrochemical analysis; nanoparticle; aptamer

Ask authors/readers for more resources

The goal of this work was to introduce a modified electrochemical sandwich model for target protein detection, exploiting antibody as the capturing probe, aptamer as the detection probe and methylene blue as the electrochemical active marker intercalating in the probing aptamer without previous labeling. With appropriate design of the sequence of the aptamer, the aptamer was successfully utilized instead of antibody for obtaining the electrochemical detection. A special immobilization interface consisting of nanogoldchitosan composite film was used to improve the conductivity and performance characteristics of the electrode. The capturing antibody was linked to the glassy carbon electrodes modified with composite film via a linker of glutaraldehyde. Differential pulse voltammetry was performed to produce the response signal. Thrombin was taken as the model target analyte to demonstrate the feasibility of proposed methodology. The sensor shows the linear response for thrombin in the range 1-60 nM with a detection limit of 0.5 nM. The proposed approach provides an alternative approach for sandwich protein assay using aptamers. (C) 2008 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available