4.7 Article

Development and characterization of an ELISA assay in PDMS microfluidic channels

Journal

SENSORS AND ACTUATORS B-CHEMICAL
Volume 72, Issue 2, Pages 129-133

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/S0925-4005(00)00640-7

Keywords

polydimethylsiloxane (PDMS); microfluidic device; immunoassay; flow-through; sensor

Ask authors/readers for more resources

This report describes the first preliminary evaluation of a heterogeneous sandwich enzyme-linked immunoassay in a PDMS microfluidic device. The PDMS devices were fabricated using replica molding against a patterned photoresist generated by photolithographic techniques. With this experimental setup, the microfluidic sensor chip was successfully used to quantify a model analyte (sheep IgM) with sensitivity down to 17 nM. The conventional blocking cocktail used in nearly all polystyrene microtiter plate-based ELISA assays failed to block the nonspecific adsorption of the analyte and secondary antibody. This report describes the successful use of surface modification chemistries and a modified blocking cocktail to reduce background due to nonspecific adsorption and to thereby increase the sensitivity of the assay. These protocols can be extended to the detection of a variety of analytes by immunoassay in PDMS microchannels. (C) 2001 Elsevier Science B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available