4.5 Article

Pressure-resisting cell for high-pressure, high-resolution nuclear magnetic resonance measurements at very high magnetic fields

Journal

REVIEW OF SCIENTIFIC INSTRUMENTS
Volume 72, Issue 2, Pages 1463-1471

Publisher

AMER INST PHYSICS
DOI: 10.1063/1.1334630

Keywords

-

Ask authors/readers for more resources

A pressure-resisting cell system has been developed for high-pressure high-resolution nuclear magnetic resonance (NMR) measurements up to a maximum pressure of 600 MPa. This cell system is capable of performing high-pressure experiments with any standard spectrometer, including modern high field NMR machines. A full description of the high-pressure NMR assembly mounted on a 750 MHz spectrometer is presented along with a detailed explanation of the procedure for preparing the pressure-resisting quartz and glass cells. (C) 2001 American Institute of Physics.